WFDB Software Package 10.7.0
(8,779 bytes)
.TH EPICMP 1 "10 July 2003" "WFDB 10.3.8" "WFDB Applications Guide"
.SH NAME
epicmp \- ANSI/AAMI-standard episode-by-episode annotation comparator
.SH SYNOPSIS
\fBepicmp -r\fR \fIrecord\fR \fB-a\fR \fIreference-annotator test-annotator\fR [ \fIoptions ... \fR ]
.SH DESCRIPTION
.PP
This program implements the VF, AF, and ST episode-by-episode
comparison algorithms specified by the current American National
Standard for ambulatory ECG analyzers (ANSI/AAMI EC38:2007). \fBepicmp\fR
is the reference implementation of these algorithms, and must be used
to obtain the episode-by-episode performance statistics cited in EC38
in order to be in compliance with the standard (see EC38, section 5.2.14).
.PP
Input to this program consists of two annotation files associated with the same
\fIrecord\fR. One of these is designated the \fIreference\fR annotation file,
the other the \fItest\fR annotation file.
.PP
\fIOptions\fR include:
.TP
\fB-A\fR \fIfile\fR
Append atrial fibrillation detection reports to the specified \fIfile\fR.
.TP
\fB-f\fR \fItime\fR
Begin the comparison at the specified \fItime\fR (default: 5 minutes after the
beginning of the record).
.TP
\fB-h\fR
Print a usage summary.
.TP
\fB-i\fR \fItime\fR
Exclude episodes shorter than \fItime\fR (default: 0 seconds) from
episode statistics.
.TP
\fB-I\fR \fItime\fR
Exclude episodes shorter than \fItime\fR (default: 0 seconds) from
episode and duration statistics.
(At most one of \fB-i\fR and \fB-I\fR may be used.)
.TP
\fB-l\fR
Write reports in line format (default: matrix format).
.TP
\fB-L\fR
Same as \fB-l\fR.
.TP
\fB-S\fR \fIfile1 file2\fR
Append ischemic ST episode detection reports to \fIfile1\fR, and ST deviation
measurements to \fIfile2\fR.
.TP
\fB-S0\fR \fIfile1 file2\fR
As for \fB-S\fR, but report on signal 0 only.
.TP
\fB-S1\fR \fIfile1 file2\fR
As for \fB-S\fR, but report on signal 1 only.
.TP
\fB-t\fR \fItime\fR
Stop the comparison at the specified \fItime\fR (default: the end of the record
if it is defined, the end of the reference annotation file otherwise; if
\fItime\fR is 0, the comparison ends when the end of either annotation file is
reached).
.TP
\fB-V\fR
Append ventricular flutter and fibrillation detection reports to the specified
\fIfile\fR.
.TP
\fB-x\fR
Exclude periods of atrial fibrillation from calculations of atrial fibrillation
positive predictivity, as required by EC38:1998 (default: include these periods,
as required by EC38:2007).
.PP
The episode and duration statistics gathered by \fBepicmp\fR are based on tallies
of overlapping episodes in the reference and test annotation files.
Duration statistics give weight to each episode or detection in
proportion to its duration.
Episode statistics give equal weight to each episode or detection,
irrespective of length; each test-annotated episode that meets the criteria
for overlap (see below) with a reference-annotated episode is counted as
a true positive. Episodes are defined as follows (see
\fI<wfdb/ecgcodes.h>\fR for definitions of annotation types):
.TP
\fIAtrial fibrillation episodes\fR
begin with a \fBRHYTHM\fR annotation, with the \fIaux\fR field containing
the text `\fB(AFIB\fR', and end with any other \fBRHYTHM\fR annotation
(or at the end of the record). Reference-marked episodes of atrial flutter
(begun by \fBRHYTHM\fR annotations with the text `\fB(AFL\fR') are excluded
from AF comparisons (i.e., the test annotator is neither penalized nor rewarded
for its treatment of atrial flutter in this context). Any amount of overlap
is sufficient to qualify a test episode as a true positive.
.TP
\fIVentricular fibrillation or flutter episodes\fR
begin with a \fBVFON\fR annotation, and end with a \fBVFOFF\fR annotation
(or at the end of the record). \fBRHYTHM\fR annotations are ignored in this
context by \fBepicmp\fR. Any amount of overlap is sufficient to qualify a test
episode as a true positive.
.TP
\fIIschemic ST episodes\fR
begin with a \fBSTCH\fR annotation, with the \fIaux\fR field containing the
text `\fB(ST\fIns\fR', and end with another \fBSTCH\fR annotation, with the
text `\fBST\fIns\fB)\fR' (or at the end of the record). Between these
annotations, the extremum (the time at which the absolute value of the ST
deviation is greatest) is marked with another \fBSTCH\fR annotation, with
the text `\fBAST\fInsm\fR'; this annotation may be omitted in the test
annotation file. In these annotations, \fIn\fR is `\fB0\fR' or
`\fB1\fR', and denotes the affected signal; \fIs\fR is `\fB+\fR' for episodes
of ST elevation, or `\fB-\fR' for episodes of ST depression; and \fIm\fR is
the ST deviation in microvolts, relative to a reference level established from
the first 30 seconds of the record. The values of \fIs\fR and \fIm\fR are not
significant for the episode comparison made by \fIepicmp\fR. When using the
\fB-S0\fR or \fB-S1\fR options, \fIn\fR must be 0 or 1 respectively; other
\fBSTCH\fR annotations are ignored. When using the \fB-S\fR option, the value
of \fIn\fR is ignored: each `\fB(ST\fIns\fR' annotation increments a counter,
and each `\fBST\fIns\fB)\fR' annotation decrements the counter; in this
context, ST episodes begin when the counter becomes positive and end when the
counter reaches zero (or at the end of the record). To qualify a test episode
as a true positive for purposes of determining ST episode sensitivity, it must
overlap at least 50% of the reference episode, or the overlap must include the
reference-marked extremum. To qualify a test episode as a true positive for
purposes of determining ST episode positive predictivity, the reference episode
must overlap at least 50% of the test episode, or the overlap must include the
test-marked extremum, if present.
.PP
The second file generated when using the `\fB-S\fR', `\fB-S0\fR', or
`\fB-S1\fR' options contains comparisons of ST deviation measurements wherever
such measurements are available in the reference annotation files. In the
existing databases, these appear only at extrema within each annotated
ischemic (or non-ischemic) ST episode, as described above.
For purposes of comparison of ST deviation measurements, test ST measurements
for each signal are read from the \fIaux\fR field of beat annotations, which
should contain text of the format `\fIm n\fR' (where \fIm\fR and \fIn\fR
are the measured ST deviations for signals 0 and 1 respectively). If these
measurements are missing from any test beat annotation, \fBepicmp\fR assumes that
they have not changed since they last appeared. \fBepicmp\fR ignores
`\fBAST\fR...' annotations in the test annotation file when making this
comparison. In the output file, any test measurements that deviate from the
reference measurements by more than 100 microvolts are tagged with an asterisk
(`\fB*\fR'). \fBplotstm\fR(1) can produce a scatter plot of these data using
this file as input.
.PP
At least one of the options `\fB-A\fR', `\fB-S\fR', `\fB-S0\fR', `\fB-S1\fR',
and `\fB-V\fR' must be used. If `\fB-\fR' is given as a \fIfile\fR argument,
reports are written on the standard output. The output generated by selecting
\fB-l\fR or \fB-L\fR includes column headings only if a \fIfile\fR other than
`\fB-\fR' is specified, and only if the specified \fIfile\fR does not already
exist. In this way, \fBepicmp\fR can be used repeatedly to build up line-format
tables for multiple records, for further processing by \fBsumstats\fR(1).
.SH ENVIRONMENT
.PP
It may be necessary to set and export the shell variable \fBWFDB\fR (see
\fBsetwfdb\fR(1)).
.SH DIAGNOSTICS
.TP
\fInon-standard comparison selected\fR
The \fB-f\fR, \fB-i\fR, \fB-I\fR, and \fB-t\fR options modify the comparison
algorithms used by \fBepicmp\fR in ways not permitted by EC38. These options
are provided for the use of developers, who may find them useful for obtaining
a more detailed understanding of algorithm errors.
.SH BUGS
.PP
Since \fBepicmp\fR performs multiple passes over its input files, it cannot be
used at the end of a pipe.
.PP
Between 1992 and 2002, this program was known as \fBepic\fR; the name
was changed to avoid conflict with a new but widely distributed IRC
chat client also named \fBepic\fR. By analogy to \fBbxb\fR, \fBmxm\fR, and
\fBrxr\fR, this program should have been called \fBexe\fR, which would
have created interesting possibilities for confusion.
.SH SEE ALSO
bxb(1), ecgeval(1), mxm(1), plotstm(1), rxr(1), setwfdb(1), sumstats(1)
.br
\fIEvaluating ECG Analyzers\fR (in the \fIWFDB Applications Guide\fR)
.br
\fIAmerican National Standard ANSI/AAMI EC38:1998, Ambulatory
Electrocardiographs\fR; available from AAMI, 1110 N Glebe Road,
Suite 220, Arlington, VA 22201 USA (http://www.aami.org/).
.SH AUTHOR
George B. Moody (george@mit.edu)
.SH SOURCE
http://www.physionet.org/physiotools/wfdb/app/epicmp.c